• Ешқандай Нәтиже Табылған Жоқ

Detector Calibration and Performance

708

8.1. Gain Calibration

709

The low level charge calibration converts raw ADC response of the electronics to photoelec-

710

tron units. It is performed in three steps: pedestal subtraction, correction for the electronics

711

non-linearity and the relative low/high gain response, and correction for the MPPC gain varia-

712

tions.

713

The pedestal, i.e. the baseline response of the MPPC and electronics without input signal,

714

as well as the MPPC gain, is measured using non zero-suppressed dark rate noise (Figure 17).

715

The pedestal peak in the dark noise spectrum is fit to a Gaussian function in each integration

716

cycle separately to account for the small variations among the cycles. The mean of the Gaussian

717

33

gives the pedestal constant used for the pedestal subtraction. The MPPC gain is measured as the

718

separation between the pedestal and the 1 p.e. peak after combining the dark noise spectra from

719

all 23 integration cycles which were corrected for individual pedestal shifts. The two peaks are

720

fit to a double Gaussian and the difference in their means is used to measure the photoelectron

721

unit in ADC values. Since the MPPC overvoltage, as well as the pedestal, is quite sensitive to

722

the temperature at a fixed bias voltage, the gain and pedestal require continuous monitoring and

723

updating.

724

ADC count

1500 160 170 180 190 200

1000 2000 3000 4000 5000

Figure 17: Typical digitized dark noise spectrum of an MPPC with a double Gaussian function fitted to the pedestal and 1 p.e. peaks.

Before converting the signal into photoelectron units, the raw ADC response needs to be

725

corrected for the non-linearity of the electronics and the relative gain difference between the

726

high and low gain response. The response of each input channel is measured using the internal

727

TFB calibration circuit as a series of 174 signal levels that covers both the high and low gain

728

dynamic range. This measurement is performed in situ when the MPPCs are powered since the

729

capacitance of the photosensors and the mini-coax cables connecting the sensors to TFBs repre-

730

sent a significant additional capacitance on the input, altering the effective electronics gain. The

731

measured response as a function of the calibration level is fit to a bi-cubic polynomial with nine

732

free parameters. This parametrization is used to correct the raw ADC values during the offline

733

calibration of the data. The bi-cubic function allows an adequate representation of the Trip-t non-

734

linearity with residuals typically smaller than a few percent (Figure 18). The electronics gain and

735

non-linearity are fairly stable, requiring only occasional checks, and therefore the constants are

736

updated only if there is a hardware change to the front-end electronics.

737

34

High Gain ADC

0 100 200 300 400 500 600 700 800

Input Charge (arbitrary units)

0 50 100 150 200 250 300 350 400 450

Figure 18: Charge versus ADC for a high-gain Trip-t channel fit to a bi-cubic function.

8.2. MIP Light Yield

738

A useful characteristic of minimum ionizing particles (MIPs) is that their energy loss is only

739

weakly dependent on their energy. Therefore, for high energy muons passing through the detec-

740

tor, the mean energy deposition per unit length is a constant. With a sample of through-going

741

muons, this constant can be determined. The first sample used was the cosmic checkout data,

742

selecting tracks with cosθ >0.8 [16] (whereθis the angle that the track makes with the z axis of

743

the detector) but this was for individual SuperPØDules only.

744

After the PØD was installed in the basket, it became possible to calibrate all PØDules with the

745

same data sample. The best sample was through-going muons from beam neutrino interactions

746

in the wall or sand and rock upstream of the PØD. After reconstruction, events were selected

747

with a single 3D track entering the front face of the PØD and exiting out the downstream end.

748

These events were analyzed to show each layer’s detection efficiency. Due to the triangular

749

design of the PØD’s scintillator bars, a normally incident MIP is most likely to pass through two

750

bars, as demonstrated in Fig. 19. However, depending on the path taken, there is a chance that one

751

bar is untouched, or that the signal is below the noise threshold cut applied by the reconstruction.

752

The results, shown in Fig. 20, indicate the probability of finding 0, 1 or 2 hits in each x or

753

y plane. The tracking efficiency is 100% for all but the first three scintillator planes, which is

754

explained by the selection criteria allowing a small number of first layer neutrino interactions

755

into the sample.

756

Figure 21 shows the summed charge deposit for the two hit sample, after calibration and

757

path correction. The plot has been fit with a Gaussian-Landau distribution, and returns a most

758

probable value of 37.9 p.e./mip/cm. This value provides a known point, which each channel of

759

35

Figure 19: Illustration of a singlet and doublet, as a MIP passes through a PØDule layer.

the PØD can be calibrated to, ensuring a constant response for the detector.

760

8.3. LIS Operation and Performance

761

The LIS system simultaneously illuminates the entire PØD and is read out at in bursts of 20

762

Hz interspersed with other trigger types. The current settings give the LIS system an effective

763

rate of 1.5Hz. The LIS system cycles through a set of ten amplitudes, each with 500 flashes,

764

taking about one hour for a complete cycle. Figure 22 shows the average ADC signal produced

765

by each of the four pulser boxes during a typical run. Each plateau corresponds to a single

766

amplitude. The sequence of amplitude was purposefully chosen to produce a clear step structure

767

in the response to enable easy visual separation of the groups from each other.

768

Besides providing the ability to quickly determine the correct functioning of all PØD pho-

769

tosensors, the LIS provides a tool to monitor the stability of the photosensor signal, shown for

770

a portion of a physics run is shown in Fig. 23. The variation over short periods of time can be

771

attributed to changes in the photosensor gain. Shifts that are different with respect to each pulser

772

can be evidence for malfunctions in the PØD readout electronics.

773

8.4. Water Target Filling and Monitoring

774

The depth sensors were found to have fluctuations of±1 mm but had a±15 mm calibration

775

offset before insertion into the PØD. This offset was reduced by using the fixed binary wet-dry

776

level sensors to provide calibration reference points in situ. We expected the water level to drop

777

36

X Layers

0 5 10 15 20 25 30 35 40

Proportion (%)

0 10 20 30 40 50 60 70 80

Y Layers

0 5 10 15 20 25 30 35 40

Proportion (%)

0 10 20 30 40 50 60 70 80

Figure 20: The PØD layer detection efficiency. The plots show the proportion of tracks with 0 (green circles), 1 (blue triangles) and 2 or more (red inverted triangles) hits in a layer, for both x and y. The small excess of 0 hits in the upstream x layer is due to neutrino interactions in the first y layer passing the cuts.

in some layers due to deflection of the plastic scintillator. As shown in Fig. 24, the largest change

778

in water level is closest to the downstream end of the PØD which is not directly supported by the

779

basket.

780

Geometry and the measured dimensions of the PØD constrain the uncertainty on the total

781

mass of water in the fiducial volume to approximately 3%. The addition of measurements from

782

the WL400 depth sensors and the external tank volume measurements reduce this uncertainty to

783

less than 1%.

784

СӘЙКЕС КЕЛЕТІН ҚҰЖАТТАР