Talanta 223 (2021) 121718

Available online 14 October 2020

0039-9140/© 2020 The Author(s). Published by Elsevier B.V. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

Review

Review: Detection and quantification of proteins in human urine

Sultan Aitekenov

a

, Abduzhappar Gaipov

b

, Rostislav Bukasov

a,*

aSchool of Sciences and Humanities, Department of Chemistry, Nazarbaev University, Nur-Sultan, Kazakhstan

bSchool of Medicine, Department of Clinical Sciences, Nazarbaev University, Nur-Sultan, Kazakhstan

A R T I C L E I N F O Keywords:

Urine proteomics Biomarkers Human serum albumin Immunoassays Mass spectrometry Fluorescence spectroscopy

A B S T R A C T

Extensive medical research showed that patients, with high protein concentration in urine, have various kinds of kidney diseases, referred to as proteinuria. Urinary protein biomarkers are useful for diagnosis of many health conditions – kidney and cardio vascular diseases, cancers, diabetes, infections. This review focuses on the instrumental quantification (electrophoresis, chromatography, immunoassays, mass spectrometry, fluorescence spectroscopy, the infrared spectroscopy, and Raman spectroscopy) of proteins (the most of all albumin) in human urine matrix. Different techniques provide unique information on what constituents of the urine are. Due to complex nature of urine, a separation step by electrophoresis or chromatography are often used for proteomics study of urine. Mass spectrometry is a powerful tool for the discovery and the analysis of biomarkers in urine, however, costs of the analysis are high, especially for quantitative analysis. Immunoassays, which often come with fluorescence detection, are major qualitative and quantitative tools in clinical analysis. While Infrared and Raman spectroscopies do not give extensive information about urine, they could become important tools for the routine clinical diagnostics of kidney problems, due to rapidness and low-cost. Thus, it is important to review all the applicable techniques and methods related to urine analysis. In this review, a brief overview of each tech- nique’s principle is introduced. Where applicable, research papers about protein determination in urine are summarized with the main figures of merits, such as the limit of detection, the detectable range, recovery and accuracy, when available.

1. Introduction 1.1. Urine composition

Urine is a readily available liquid for the medical diagnosis of pa- tients. Urine tests are non-invasive procedures that involve no pain or discomfort for patients to determine problems with kidney function since extensive medical research showed that patients, with high protein concentration in urine, have various kinds of illnesses of the kidney, referred to as proteinuria [1]. Thus the precise and simple determination of urinary concentrations of total protein is important for diagnostic purposes. Since urine is a complex matrix, it poses challenges for the analytical determination of proteins and other constituents. Among the reasons that make urine challenging for researchers to analyze are: urine matrix is complex, it consists of various inorganic and organic com- pounds, from low-molar mass molecules to polymers; urine could contain cells, such as blood cells, or bacteria, which changes the composition of urine in time rapidly; an analytical method for diagnosis

of proteinuria should cover protein presence in urine in a wide range from 0.01 mg/ml to 10 mg/ml. Also, it is important to note that the concentration of protein in urine taken from patients can vary widely depending on dieting, exercising, and time of the day a patient urinated.

It is widely accepted that analysis of urea taken throughout a day is the best representation of any illnesses in the human body if any. The so-called – the urine 24-h volume test, that measures the amount of urine the human body produces in a day. Another option is to determine creatinine to protein ratio in urea since creatinine to protein ratio is positively correlated with 24-h volume test for quantifying proteinuria [2,3]. The noninvasive collection of samples and wide range of diag- nostic targets found in urine makes urinalysis well suited for point-of-care (PoC) monitoring applications, in which testing is done at the time and place of patient care [4]. Table 1 shows constituents in urine taken from the urine 24-h volume test.

In healthy individuals, Tamm–Horsfall protein (also known as uro- modulin) is the most abundant protein in urine (50%), followed by al- bumin (20%) and immunoglobulin (5%) [6]. Other constituents in the

* Corresponding author.

E-mail address: [email protected] (R. Bukasov).

Contents lists available at ScienceDirect

Talanta

journal homepage: www.elsevier.com/locate/talanta

https://doi.org/10.1016/j.talanta.2020.121718

Received 9 July 2020; Received in revised form 23 September 2020; Accepted 26 September 2020

(2)

urine of either sick or healthy patients include, but not limited to:

glucose; low abundant proteins, such as bilirubin, and urobilinogen;

cells, such as Erythrocyte, Leukocyte, and other cells; bacteria [7].

1.2. Urine as a diagnostic tool

The urine of a healthy individual contains up to 150 mg of protein in total measured throughout a day, of which approximately 20 mg is al- bumin (human serum albumin) [1]. Albumin excretion of 30–300 mg a day, which is called microalbuminuria, is an early and sensitive marker of diabetic nephropathy [8], cardiovascular and renal disease [9]. The 15–30 mg/L albumin concentration is a critical value that could indicate kidney problems when it is repeatedly exceeded [10]. Though one of the main proteins in urine is albumin, there are thousands of other types of proteins. Furthermore, the removal of albumin from the urine helps to identify the low abundant proteins [11]. Marimuthu et al. by using high-resolution Fourier transform mass spectrometry were able to identify 1823 proteins in the urine of healthy subjects [12]. Using mass spectroscopy and sub-fractionating normal urine by successive steps (vesicle separation, CPLL, and solvent treatments) Santucci et al. were able to identify 3429 individual proteins [13]. All those discovered proteins could be potential biomarkers for diseases. In a review paper by R¨othlisberger et al. it was stated that besides albumin, other proteins such as CD14, hh-FABP, BNP/NT-proBNP, NGAL, ORM1 are potential biomarkers of cardiovascular disease [14]. Another biomarker in human urine, Bence Jones protein (BJP), has an important diagnosis and prognosis value for multiple myeloma, cancer formed in a white blood cell called plasma cell [15]. Other types of chemical substances in urine can be important biomarkers as well, for example, urine microRNAs have the potential to be a valid marker for bladder cancer detection [16].

Proteinuria is the main clinical presentation of glomerular diseases (the glomerulus are complex capillary set that are located in the neph- rons – renal cells) [17]. The glomerular diseases can be primarily when the disease caused by kidney diseases (such as glomerulonephritis), or secondary, when kidney glomerulus becomes target organ affected by different diseases such as diabetes and cardiovascular diseases, auto- immune and inflammatory disorders, amyloidosis and neoplasms, can- cer and many other including genetic disorders [17,18]. The levels of proteinuria, which usually measures in clinical settings, may vary depending on severity of disease and glomerular injury, despite any cause of disease. Any detectable proteinuria in the urine between 30 and 300 mg/24h called albuminuria (or microalbuminuria), above the 300 mg/24h – proteinuria. Depending the amount of proteinuria they considered to named nephritic range proteinuria below 3.5 g/24h and nephrotic range proteinuria when protein loss excess the 3.5 g/24h (heavy proteinuria). The most of the cases, proteinuria caused by car- diovascular diseases limited with nephritic range proteinuria, whereas in renal diseases and cancer the proteinuria may reach nephrotic range proteinuria [19,20]. Since urine is in direct contact with kidneys, analysis of urine is important for diagnosis of renal diseases, such as chronic kidney disease. Several viral infections such as Epstein-Barr virus, hepatitis B and C viruses, herpes zoster, hantavirus, human im- munodeficiency virus, dengue fever, COVID-19 and many others also may lead to proteinuria [21,22]. The potential mechanisms of

proteinuria are related to primary affect to glomerulus and/or secondary to autoimmune response to infection [23,24]. Chronic kidney disease (CKD) is a major public health problem. Albuminuria, urinary sediment abnormality and other markers of kidney damage are criteria of CKD, according to international guidelines [25]. Fassett et al. summarized biomarkers for CKD in urine – cystatin C, β-trace protein, NGAL, KIM-1, NAG and many others [26]. A more recent article on this topic is dis- cussed in Ref. [27].

In the United States, it was estimated in 2003 that 11% of the adult population (aged 20 or older) has chronic kidney disease [28]. In a re- view paper about the prevalence of chronic kidney disease (CKD) by Zhang et al., the authors studied relevant data across America, Europe, Asia, Australia [29]. They reported that the median prevalence of CKD was 7.2% in persons aged 30 years or older. In persons aged 64 years or older prevalence of CKD varied from 23.4% to 35.8%. The authors concluded that worldwide, CKD is becoming a common disease in the general population. In a more recent paper, Wei et al. studied data on kidney damage among elderly people in Wuhan, China [30]. The age-standardized prevalence of kidney damage decreased renal function and proteinuria was 17.2, 13.5, and 5.3%. In the US, patients aged 66 or older have a mortality rate of 111.2 per 1000 due to CKD, compared to 45.2 per 100 due to non-CKD reasons as of 2014 [31]. It remains among the few growing causes of mortality which made CKD the 13th leading cause of death in 2013 [32].

Other renal diseases include diabetic nephropathy, autosomal dom- inal polycystic kidney disease, paediatric renal disease, acute kidney injury, and renal transplant rejection [33]. Besides renal diseases, urine can also be a useful source of information related to cancers and non-renal diseases. Theodorescu et al. found biomarkers in urine related to urothelial carcinoma, also called transitional cell carcinoma, by the means of capillary electrophoresis coupled with mass spectroscopy [34].

It is the most common type of bladder cancer. Bhasin et al. developed a bioresistor device to detect bladder cancer marker DJ-1 in urine at concenctration of 10 pM in 1 min [35]. Zhang et al. in their review summarized information about cancer biomarkers in urine [36]. In another work related to non-renal disease, Zimmerli et al. investigated urine samples of patients with coronary artery disease [37]. Rossing et al. identified potential biomarkers for diabetes [38]. Raja et al.

investigated into the potential of albuminuria as a biomarker of diabetic complications [39]. Moreover, researchers could diagnose patients with viruses from their urine samples. Yang et al. developed an immunoassay cassette with a handheld reader for HIV urine testing in point-of-care diagnostics [40]. Robles et al. analyzed viruses present in urine from patients with interstitial cystitis [41]. Other viruses in urine described in literature include – Zika virus [42], hepatitis C [43], human papiloma virus [44]. Niedrig et al. wrote a review about usefulness of saliva and urine for the diagnosis of emerging viruses [45]. They concluded that it is important to perform an investigation using non-invasive approaches, such as urine can provide, for the diagnostic of emerging viral diseases.

1.3. Clinical urine tests

Urinalysis is an abbreviation for clinical urine tests for diagnostic purposes. A urinalysis (UA) is one of the most common methods of medical diagnosis. There are three basic components to urinalysis [46]:

• Gross/physical examination targets parameters that can be measured or quantified with the naked eye (or other senses), including volume, color, transparency, odor, and specific gravity.

• Microscopic examination. The numbers and types of cells and/or material such as urinary casts can yield a great detail of information and may suggest a specific diagnosis.

• Chemical examination of urine measures quantitatively and quali- tatively for pH, blood, nitrite, protein, glucose, ketones, bilirubin, urobilinogen, ascorbic acid.

Table 1

Composition of urine (averages of selected components in 24-h collection test) [5].

Component Average weight (mg)

Water 1,200,000.0

Urea 24,000.0

Creatinine 1335.0

Uric acid 505.0

Albumin 90.0

(3)

Conditions for storage of urine samples is an important consideration for chemical analysis, since undesired changes in unpreserved urine occur, such as a decrease of concentration of glucose due to consumption of it by cells or bacteria; a decrease of concentration of bilirubin due to photo-oxidation and etc [47]. The most common form of preservation of urine samples is refrigeration, also chemicals can be utilized too. Remer et al. suggest that high long-term stability and measurement validity for numerous clinical chemistry parameters stored at − 22 C without the addition of preservative for human urine can be achieved [48]. Specif- ically for free-light chains monoclonal immunoglobulin, Pieri et al.

suggest to store urine sample at +4 C, if analysis to be made within 72 h, otherwise for longer periods, the refrigeration should be done at

− 80 C [49]. Moyle et al. found that collection and storage of canine urine samples in clean homopolymer polypropylene (HP), propylene copolymer (PC), or glass containers at 24 C for 4 h, 4 C for 12 h or

− 20 C for 72 h is unlikely to result in clinically relevant decreases in measured protein to creatine values [50].

Historically, qualitative or semi-quantitative screening tests for urine protein relied on protein precipitation techniques. Proteins denature upon exposure to extremes of pH or temperature, and the most visible evidence of this is a decrease in solubility. In clinical laboratories, sul- fosalicylic acid at room temperature may be used to detect urine protein [47]. This protein precipitation method detects all proteins—albumin and globulins. The analytical signal here is turbidity, which is the cloudiness or haziness of a fluid. This method can not detect protein below 0.05 mg/ml. Modern commercial reagent strips are available for routine protein screening use change in color due to acid-base reaction or formation of complexes between urine constituents and reagent strips material. Some of those reactions are described and compared in Ref. [51]: absorbance at 280 nm, which monitors tyrosine and trypto- phan in protein (UV absorbance); the method of Lowry et al. based on complex formation between tyrosine, tryptophan or cysteine and heavy metals using Folin phenol reagent in the presence of copper ions under alkaline conditions (Lowry method) [52]; the Bradford method based on protein dye-binding using Coomassie brilliant blue (CBB method) [53];

the method of Smith et al. based on the formation of a purple complex between bicinchoninic acid and copper ions under alkaline conditions (BCA method) [54]; the method of Watanabe et al. based on protein-dye binding using pyrogallol redmolybdate reagent (PRM method) [55].

To date, different diagnostics methods in clinical settings can be used to ascertain urinary protein and albumin. Urine dipstick test, technique with acidic buffer precipitation and immuno-electrophoresis are commonly used diagnostic methods for urinary total protein quantifi- cation [56]. The protein-specific dipstick and immunochemical tech- niques are the fastest and cheapest way to determine proteins in the urine for diagnostic purposes. Modern automated urine analyzers are based on similar principles and are widely used in clinical routine tests [57]. A urine analyzer is a device used in the clinical setting to perform automatic urine testing. The units can detect and quantify a number of analytes including bilirubin, protein, glucose, and red blood cells. Many models contain urine strip readers, a type of reflectance photometer that can process several hundred strips per hour. Some analyzers can perform up to 240 tests an hour. Prices of urine analysis are relatively low, but they depend on a location/country and they are available from the most clinical laboratories The high-performance liquid chromatography mostly used for scientific purpose due to its relatively high cost, though relatively inexpensive versions of mass spectrometers are now available [58].

Prices of urine analysis are relatively low, but they depend on a location/country and they are available from the most clinical labora- tories The high-performance liquid chromatography mostly used for scientific purpose due to its relatively high cost, though relatively inexpensive versions of mass spectrometers are now available [58].

1.4. Focus of this review

Several review papers are focused on urine proteomics. A review by Albalat et al. focuses on urinary proteins as potential biomarkers for mainly urine diseases, and capillary electrophoresis coupled mass spectroscopy as an instrumental method [33]. The review also gives information on biomarkers on non-renal diseases obtained from the urine. Kalantari et al. state that mass spectrometry as a detection tech- nique is the most common [59]. The paper discusses technical aspects of urinary proteomics, proteomic technologies, and their advantage and disadvantages. As of 2015, several recent experiments are presented there which applied urinary proteome for biomarker discovery in renal diseases including diabetic nephropathy, immunoglobulin A (IgA) ne- phropathy, focal segmental glomerulosclerosis, lupus nephritis, mem- branous nephropathy, and acute kidney injury. Decramer et al. in their review paper focus on mass spectrometry-based urinary protein and peptide profiling [60]. The advantages and disadvantages of different mass spectroscopy methods are compared. Applications of urinary proteome analysis to urogenital and non-urogenital diseases are also discussed there.

This review, besides covering mass spectrometry, holds chapters discussing fluorescence spectroscopy, immunoassay, infrared, and Raman spectroscopy that was not given much attention compared from the reviews on urinalysis mentioned in the previous paragraph. This review also covers separation techniques – electrophoresis and chro- matography. For each technique, a brief overview of the technique’s principle is introduced. Where applicable, research papers about protein determination in urine are summarized with the main figures of merits, such as the limit of detection, the detectable range, recovery, and ac- curacy. Particularly, electrophoresis, chromatography, immunoassay, and fluorescence spectroscopy chapters are followed by summary tables with analytical parameters taken from individual research papers. It should be noted that not all analytical parameters are given in every paper, for example, sometimes recovery is unknown, and thus is not included in the following tables. Since the volumes used for all tech- niques are 100–1000 times smaller than a typical urine probe available from one patient for analysis, usually 50–100 mL, urine volume is not introduced in the tables. For instance for proteomics study, a typical sample size for high performance liquid chromatography (HPLC) ranges from few microliters to 0.1 mL [61]. A commercial immunoassay test kits developed by Valle et al. require 10 μL of urine [62]. This review ends with a paragraph about urinary protein biomarkers. Most of those biomarkers were identified by mass spectrometry or immunoassay methods. A summary table of experimental papers on urinary protein biomarkers with related diseases or health conditions is given.

2. Separation techniques 2.1. Electrophoresis

Electrophoresis is an electrokinetic process, with a long history of development, which separates charged particles (macro-molecules, such as proteins) in a fluid using a field of electrical charge [63]. Particles could be separated based on their charge or their mass. The electro- phoretic method is a widely used method in a modern research labo- ratory, but not so much in a clinical laboratory. Electrophoresis is commonly used for proteins, peptides and nucleic acid analysis [64]. It could be applied as a qualitative or quantitative analytical technique on its own if a reference substance is given, or it could be used as a sepa- ration technique for a sample to be further examined by other tech- niques, such as mass spectroscopy, fluorescence spectroscopy and etc.

Aguzzie et al. successfully used the electrophoretic technique to detect Bence-Jones protein (BJP) in human urine with the limit of detection LOD of 1 mg/L to BJP, and LOD of 3 mg/L to other types of protein [65]. These results were obtained by immunofixation using Dako antisera on cellulose nitrate followed by further staining with gold.

(4)

The authors claim that their technique does not require particular skill and is cheap. Such low levels of LOD allowed them to use uncon- centrated urine samples, thereby eliminating the problem of the loss of low molecular weight proteins. Usually, though it is often necessary to concentrate the urine before electrophoresis because of the low urinary protein level [6]. This is a time-consuming process. Machii et al. used cellulose acetate membrane electrophoresis followed by colloidal silver staining on the urine of healthy subjects with success [6]. However, when they used silver staining on an agarose gel, urinary protein in healthy subjects was not detected. They estimated that urine has to be concentrated 10 times, furthermore, the pattern of the urinary fraction will change after the concentration procedure is conducted. Giovannoli et al. developed a noncompetitive capillary electrophoresis immuno- assay with laser-induced fluorescence detection was used to determine human serum albumin (HSA) in buffer and urine [66]. The injection of an excess of labeled antibody off-line incubated with the analyte allows the surface capture of the free antibody and consequently the immu- nocomplex detection. In buffer, authors were able to achieve the limit of detection LOD of 0.60 mg/L, with the quantitative range up to 6.6 mg/L.

Recovery in the urine matrix was dependent upon sample dilution and HSA added and was as low as 91 ±8%.

One of the variations of the electrophoretic method is called gel electrophoresis, where a gel is used as an anticonvective medium.

Particularly polyacrylamide gel PAGE is used for separating proteins due to the uniform pore size provided by the polyacrylamide. Jia et al. used PAGE gel electrophoresis with sodium dodecyl sulfate (SDS-PAGE) to show a good correlation (R2 =0.825) with the pyrogallol red-molybdate (PRM) assay technique used in clinical laboratories for urine samples [67]. Sodium dodecyl sulfate (SDS) coats proteins with a negative charge so that proteins are separated due to their size, not charge. The authors estimated that their method can detect protein as low as 5 mg/L.

Tran et al. showed that the use of tube gel electrophoresis for protein separation was further expanded with the invention of gel-eluted liquid

fraction entrapment electrophoresis (GELFrEE) [68]; shown in Fig. 1.

Bessonova et al. in their review paper compared capillary electro- phoresis with five on-line preconcentration techniques, including field- amplified sample stacking (FASS), head-column field-amplified sample stacking (HC-FASS), stacking with a polymer solution, dynamic pH junction and large-volume sample stacking (LVSS) with reversed po- larity for various analytes including protein in the urine [69]. The use of these preconcentration techniques makes it possible to increase sensi- tivity in the determination of analytes by a factor of 100 or higher. In another work by Bessonova et al., they found that the lowest LOD was obtained by the LVSS approach, and it was 15 mg/L with UV-detection [70].

Electrophoresis is often used to be followed by another method, such as mass spectroscopy. Wittke et al. used the on-line coupling of capillary electrophoresis (CE) with electrospray mass spectroscopy to distinguish the difference in polypeptide patterns in the urine of healthy individuals and patients with kidney disease [71]. Kiprijanovska et al. used two dimensional get electrophoresis coupled with mass spectroscopy (2D PAGE – MS) for mapping and identification purposes of the urine pro- teome [72]. For the same purposes, Nakayama et al. used cellulose ac- etate membrane electrophoresis coupled with liquid chromatography mass spectroscopy LC-MS [73].

2.2. Chromatography

Another widely used separation technique in chemistry is chroma- tography. In chromatography, a sample is dissolved in a fluid called the mobile phase, which carries it through a structure holding another material called the stationary phase. The various constituents of the mixture travel at different speeds, causing them to separate due to dif- ferential affinities (strength of adhesion, as an example) of the various components of the analyte towards the stationary and mobile phase results. Then resulting constituents can be further analyzed by various

Fig. 1.Diagram of the GELFrEE device [68]. A gel column is utilized to achieve electrophoretic separation of proteins, analogous to SDS–PAGE, which are then eluted into the liquid-phase for manual collection. The fractionation can then be visualized by running a portion of the fractions on a SDS–PAGE gel. Reprinted from Refs. [95].

(5)

techniques, such as mass spectroscopy, UV-spectroscopy, fluorescence detection, etc. Depending on the mobile phase methods in chromatog- raphy could be classified as gas-chromatography (GC) [74], and liquid-chromatography (LC) [75]. Typically, proteomic sample size ranges from few microliters to 0.1 mL [61.

Ishida et al. quantitatively determined human serum albumin (HSA) in human plasma and urine by utilizing post-column high-performance liquid chromatography (HPLC) with fluorescence detection [76].

Particularly, they used 8-anilino-1-naphthalenesulfonic acid as a fluo- rophore to bind HSA. The sample size was 1 ml for urine, and the retention time was around 20 min. A linear relationship was observed between fluorescence peak and the amounts of HSA in both plasma and urine, as for the latter matrix, a linear relationship was observed up to 400 mg/L, with the correlation coefficient being 0.998. The LOD for HSA was 0.2 mg/L, while the recovery was 94.5 ±3.8%. The same authors also compared their method with the immunological nephelo- metric method and obtained the correlation coefficient of 0.990 [77].

Often researchers compare their method with the immunoassay tech- nique that is used in clinical practice. Brinkman et al. in their experi- ments using the immunonephelometric approach and HPLC determined that HPLC reveals higher values of HSA, especially in the lower con- centration range, resulting in a higher prevalence of microalbuminuria [78]. Nephelometry only measures immunoreactive albumin, whereas HPLC measures both immunoreactive and immunoreactive intact al- bumin, as was found by Comper et al. with the HPLC method [79]. The authors also compared HPLC with radioimmunoassay (RIA), immuno- nephelometry (IN), and two different methods of immunoturbidimetry (IT).

Owen et al. used a new urinary albumin assay that uses size- exclusion high-performance liquid chromatography with UV-detection [80]. The limit of detection (LOD) of the assay based on a signal/noise ratio of 5 was 3.4 mg/L. The linearity of the HPLC assay was tested in a range of albumin concentrations from 4.3 to 240 mg/L. The maximum deviation from a mean recovery of 100% was 4.6% at a concentration of 66 mg/L. Total imprecision was less than 10% from 16 to 206 mg/L.

They also conducted experiments to compare their approach with immunoturbidimetric (ITA) assay, which showed positive proportional bias, which decreased with increasing concentrations of albumin. Singh et al. used liquid chromatography with quadrupole time-of-flight mass spectrometry (Q-TOF, LC-MS) method for the quantification of urinary albumin using a15N-isotopically labeled bovine serum albumin as an internal standard in human urine matrix [81]. The limit of detection (LOD) was 4.84 mg/L, while the limit of quantification was 10.5 mg/L.

Multiple calibration curves ranging from 4 to 625 mg/L were linear and were reproducible with the coefficient of determination (R2) of 0.999.

Micellar electrokinetic capillary chromatography (MEKC) is a hybrid method that combines chromatographic and electrophoretic separation principles, extends the applicability of capillary electrophoretic methods to neutral analytes [82]. In micellar electrokinetic capillary chroma- tography, surfactants are added to the buffer solution in a concentration above their critical micellar concentrations, consequently, micelles are formed; micelles that undergo electrophoretic migration like any other charged particle. Glavac et al. utilized MEKC combined with UV-detection of urine samples of patients diagnosed with proteinuria without any sample pre-treatment before analysis [83]. They were able to simultaneously identify albumin (HSA), haemoglobin (HGB), myoglobin (MYO). For all three analytes, linear correlations were ob- tained with the correlation coefficient (R2) being not less than 0.99.

Particularly, for HSA, the limit of detection (LOD) was 0.115 mg/L, the limit of quantification (LOQ) was 0.383 mg/L, a linear range between 1 and 300 mg/L and the recovery was 101.9 ±4.5%. In another work by Wu et al., a MEKC method combining field-amplified sample injection (FASI) with UV-detection has been developed for the analysis of albumin (HSA) and transferrin (TRF) in human urine [84]. Field-amplified stacking or injection is an on-column (on-line) preconcentration tech- nique which is based on the different conductivity between two solution

plugs. For TRF, the authors obtained the following analytical parameters of the method: LOD 0.31 mg/L, linear range 1.00–100 mg/L with R2 0.994. For HSA: LOD 0.14 mg/L, linear range 0.50–100 mg/L with R2 0.999. The recoveries of TRF and HSA were 91.1–101.7%.

Jaffuel et al. performed Aquaporin-2 (AQP2) nephrotoxicity biomarker measurement by using liquid chromatography–multiple re- action monitoring cubed mass spectrometry assay (MRM mode in LC- MS/MS) [85]. AQPs regulate numerous downstream effector signaling molecules that promote cancer development and progression [86]. In numerous cancer types, AQP expression has shown a correlation with tumor stage and prognosis. Linearity is observed within the concentra- tion range 0.5⋅103-50⋅103 mg/L, intra and inter-assay precision ranged from 9 to 35% at the lower limit of quantification (LLOQ), and accuracy from 94 to 114%.

Overall, due to complex composition of biofluids, including the urine, liquid chromatog-raphy in particular HPLC, which is frequently coupled with MS for detection, remains a common method in bio- analytical laboratories to separate complex mixtures. A review by Novakova et al. states that in order to decrease time for an analysis without compromising resolution and separation efficiency, three main approaches in HPLC exist – the use of monolith columns, LC at high pressures and temperatures, and HPLC at ultra-high pressures [87]. By employing those approaches, time for analysis of biofluids could be done within few minutes up to half an hourhe Table 2 below shows a sum- mary table of quantitative methods for application of electrophoresis and chromatography for protein determination in the human urine matrix.

3. Detection techniques 3.1. Mass spectrometry

Mass spectrometry is a powerful analytical technique that measures the mass-to-charge ratio of ions. Firstly, molecules in a sample must be ionized, which causes a large molecule to fragment into smaller ions;

then those ions are separated by a magnetic field depending on their mass to charge ratio. The results are typically presented as a mass spectrum, a plot of intensity as a function of the mass-to-charge ratio.

Usually, mass spectrometry is utilized after pre-treatment of a sample by electrophoresis or chromatography.

Mass spectrometry is a widely used technique for proteomics studies.

The proteome is the entire set of proteins that is produced or modified by an organism. Research on urine proteomics is important for the identi- fication purposes of reliable biomarkers that help the diagnosis of dis- ease [90]. In a review by Kalantari et al. on urine proteomics is stated that common techniques for urinary proteome analysis are two-dimensional gel electrophoresis followed by mass spectrometry (2DE-MS), liquid chromatography coupled to mass spectrometry (LC-MS), surface-enhanced laser desorption/ionization coupled to mass spectrometry (SELDI-TOF), and capillary electrophoresis coupled to mass spectrometry (CE-MS) [59]. By using mass spectroscopy protein constituents of urine were identified. Also reviews by Albalat et al. and Decramer et al. focus on mass spectrometry-based urinary protein and peptide profiling [33,60]. The advantages and disadvantages of different mass spectrometry methods are compared. As was mentioned in the introductory part of this review, Marimuthu et al. by using high-resolution Fourier transform mass spectrometry were able to identify 1823 proteins in the urine of healthy subjects [12]. Santuccie et al. by using mass spectrometry and sub-fractionating normal urine by successive steps (vesicle separation, CPLL, and solvent treatments) were able to identify 3429 individual proteins [13].

Two ionization techniques in mass spectrometry are commonly used for protein studies: electrospray ionization (ESI) and matrix-assisted laser desorption/ionization (MALDI) [91]. In the ESI approach, ions are produced in an electrospray in which a high voltage is applied to a liquid to create an aerosol. ESI is different from other ionization

(6)

processes (e.g. MALDI) since it may produce multiple-charged ions, effectively extending the mass range of the analyzer to accommodate the kDa-MDa orders of the magnitude observed in proteins and their asso- ciated polypeptide fragments [92]. In MALDI, a sample is dispersed in a certain solid matrix (e.g. aromatic compounds), which then irradiated by the laser that causes electronic excitation of molecules in the sample matrix [93]. Both methods ESI and MALDI produce relatively high molecular ions due to mild fragmentation of those approaches. It should be noted a specific method of MS applied to biomolecules, which is called tandem mass spectrometry, also known as MS/MS. In MS/MS two or more mass analyzers are coupled together using an additional reac- tion step to increase their abilities to analyze samples. In general, pro- tein identification in mass spectroscopy is performed using either bottom-up or top-down approaches. Bottom-up protein identification methods rely on enzymatic digestion to break down proteins into smaller peptides that are easier to ionize and fragment, resulting in higher sequence coverage [94]. However, in top-down methods, intact proteins are injected into the mass spectrometer and subjected to frag- mentation without pre-treatment [95]. Aside from better tracking of protein modifications, an advantage of top-down protein sequencing is that it complements imaging experiments by enabling mass measure- ment of the intact protein that relates more directly to the MALDI IMS

generated signals [96].

Analysis of a mass-spectrum, a plot of intensity as a function of the mass-to-charge ratio, could be a challenge itself, especially for big molecules like proteins, since they can be divided in many fragments resulting in a complex spectrum. Usually, the mass-spectrum is compared by software with the existing database. Several computa- tionally feasible approaches have emerged for performing protein inference from such data [97].

Mass spectrometry in itself does not allow for quantitative analysis due to the different physicochemical properties of different peptides and proteins [98–100]. However quantification in MS is possible if an in- ternal standard is used, for such there are approaches based on labeling with stable isotopes (such as N-15, O-18), involving artificial labeling of peptides and proteins. On the other hand, there are label-free ap- proaches, in which the samples retain their native isotope composition and are compared between separate measurements. In general, they could be classified into two categories: methods that involve comparing peptide signals at the level of LC-MS analysis; methods that involve counting the number of identified peptides or acquired fragment spectra. Also, it should be noted that label-based methods are relatively laborious while label-free methods are less accurate [98]. An example of quantitative protein determination, in this case – albumin, in the urine Table 2

Summary table of quantitative methods for application of electrophoresis and chromatography for protein determination in the human urine matrix (unless otherwise stated in the method section). CV (coefficient of variation) parameters are taken from maximum readings, for Recovery – the range is given.

Method Analyte The limit of

detection, LOD

Range Analytical parameters Experimental parameters Reference

Electrophoretic with

immunofixation BJP Bence Jones

Proteins 1 mg/L NA NA Migration time 33 min and

43 min depending on buffer Voltage 220V

4 Aguzzi et al., 1993 [65]

Electrophoretic with

immunofixation Transferrin, retinol binding protein, b2- microglobulin

3 mg/L (for

all) NA NA Migration time 33 min and

43 min depending on buffer Voltage 220V

5 Aguzzi et al., 1993 [65]

HPLC high performance liquid chromatography with fluorescence detection

HSA 0.2 mg/L 0.2400 mg/L

(linear, R2 =0.998) Recovery 94.5 ±3.8% Retention time for HSA 16.0 min Fluorescence enhancement factor 1286 TSK gel G2000 SWXL column

6 Ishida et al., 1996 [76]

HPLC with UV-detection HSA 3.4 mg/L 4.3–240 mg/L

(linear) Recovery 100 ±4.6% at

66 mg/L Zorbax Bio Series GF-250

column 7 Owen et al.,

2005 [80]

CZE capillary zone electrophoresis with preconcentration techniques

HSA 15 mg/L 15–1000 mg/L

(linear, R2 = 0.9840)

Sensitivity enhancement factor by large volume sample stacking (LVSS) 67

Voltage 15/25 kV 8 Bessonova et al., 2007 [70]

Capillary electrophoresis immunoassay with fluorescence detection (the buffer matrix)

HSA 0.6 mg/L 0.6–6.67 mg/L

(detectable) Recovery 91–160% Voltage 12–15 kV 9 Giovannoli et al., 2007 [66]

Q-TOF, LC-MS using a N-15

isotope HSA 4.84 mg/L 4625 mg/L

(linear, R2 =0.999) LOQ 10.5 mg/L CV intra assay 12.6% CV inter assay 12.2%

Cone voltage for MS 4080 V 10 Singh et al., 2007 [81]

FASI-MEKC field amplified

sample injection MEKC HSA 0.14 mg/L 0.50–100 mg/L

(linear, R2 =0.999) Recovery 91.1–101.7% Retention time for HSA around 10–12 min Voltage for MEKC 20 kV Sample injection volatage 10 kV

11 Wu et al., 2009 [84]

FASI-MEKC TRF transferrin 0.31 mg/L 1.00–100 mg/L

(linear, R2 =0.994) Recovery 91.1–101.7% Retention time for TRF around 10–12 min Voltage for MEKC 20 kV Sample injection voltage 10 kV

12 Wu et al., 2009 [84]

LC–MS/MS (MRM mode) AQP2 aquaporin-2 5⋅104 mg/L 5⋅104 -5⋅102 mg/

L (linear) Accuracy from 94 to 114% CV intra assay 9%

CV inter assay 35%

Retention time around 5 min

a Symmetry™ C18 column 13 Jaffuel et al., 2013 [85]

Reversed-phase HPLC with

fluorescence detection Tamm–Horsfall

protein 0.35 mg/L 4.5–90.0 mg/L

(linear, R2 =0.999) Recovery 100.0–104.2%

CV intraday 1.73–2.77%

CV interday 2.50–5.35%

COSMOSIL 5C18-MS-II

Column 14 Akimoto

et al., 2015 [88]

SDS-PAGE sodium dodecyl sulfate–polyacrylamide gel electrophoresis

HSA 5 mg/L NA CV 21.5% Electrophoresis time 5 min

Staining time 15 min 15 Jia et al., 2019 [67]

Paper-based analytical device - ion concentration polarization PAD-ICP

HSA 10 mg/L 50350 mg/L

(linear, R2 =0.994) 10–500 mg/L (detectable)

Recovery 93108% CV

11% 16 Gao et al.,

2019 [89]

(7)

matrix by MS methods is described in a paper by Singh et al. [81]. They used N-15 isotopically labeled albumin as an internal standard. The authors achieved the limit of detection (LOD) of 4.84 mg/L, the limit of quantification of 10.5 mg/L with the linear range being from 4 to 625 mg/L, R2 =0.999. Chen et al. used isotope-labeled peptides as internal standards for quantification of 63 proteins in human urine by multiple reaction monitoring-based mass spectrometry for the discovery of po- tential bladder cancer biomarkers [101]. Bachmann et al. experimen- tally compared routine measurements of urinary albumin to isotope dilution tandem mass spectrometry (IDMS) [102]. They reported that median differences between the largest positive and negative biases versus IDMS were 45%, 37%, and 42% in the concentration intervals of 12–30 mg/L, 31–200 mg/L, and 201–1064 mg/L, respectively. Lieske et al. reported their current status on a developing reference method for quantification of urinary albumin by liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) [103]. They used three different labeled peptides as internal standards to cover the range of clinical values (5–1270 mg/L). Recovery was between 97% and 108% with an average of 103%. In conclusion, quantitative mass spectroscopy assays for peptides and proteins are difficult to implement, and the future will see standardization of MS validation and quality control requirements.

Overall, mass spectrometry coupled with liquid chromatography is a powerful tool for the discovery and the analysis of biomarkers of dis- eases from biofluid samples, which includes proteomics study of urine as well [104]. Many of urinary proteome biomarkers were discovered by this technique. However, quantitative protein determination represents a challenge by following this approach [105]. Thus, mass spectroscopy could be used for biomarker discovery but not for clinical applications [60]. Another consideration for an analytical technique is the cost of analysis. Practical usage of LC-MS is limited by factors such as instru- ment size, cost, ease of operation, though, low cost compact mass spectroscopy detectors coupled with cinematographic separation are being developed, their costs equal to around $50000 [58].

3.2. Immunoassay

Immunoassay is a widely used technique in clinical and research laboratories to determine qualitatively and quantitatively macromole- cules (usually proteins) generated from organisms [106]. Immunoassay is based on a reaction between an antigen and an antibody, which is found in the adaptive immune system of vertebrates, that includes humans [107]. The reaction is carried out in-vitro (not in a living or- ganism), and the result of it is the formation of a complex antibody-antigen formed by spatial complementary, like lock and key, thus such interactions are highly selective. Antibody-antigen complexes are bounded by weak forces such as electrostatic forces, hydrogen bonds, hydrophobic interactions, and van der Waals forces; which means that such bindings are reversible. The specific piece of the antigen to which an antibody binds is called the epitope. Usually, in analytical papers, the analyte is the antigen while the reagent is the antibody.

Immunoassay methods have been widely used in many areas of phar- maceutical analysis such as diagnosis of diseases, therapeutic drug monitoring, clinical pharmacokinetics, and bioequivalence studies in drug discovery and pharmaceutical industries [108]. Conventional immuno-chemical based urinary albumin assays for diabetic patients as well as nondiabetic patients with renal disease, including those who may suffer hypertension and cardiovascular disease, now amount to 100 million assays per year worldwide as of 2004 [79]. The widespread immunoassay methods in the pharmaceutical analysis is attributed to their inherent specificity, high-throughput, and high sensitivity for the analysis of a wide range of analytes in biological samples as well the relatively low cost of the instruments, tools, or the reagents [109].

Based on signal detection immunoassays could be classified as follows:

Immunoassay experiments could be classified into types – competi- tive and non-competitive [106,109]. The competitive immunoassay

relies on the competition between the antigen of interest (the analyte) and a constant amount of a similar but labeled antigen for a limited amount of specific antibody. When these immunoanalytical reagents are mixed and incubated, the analyte is bound to the antibody-forming an immune complex. This complex is separated from the unbound reagent fraction by physical or chemical separation technique, though some- times separation is not required. Based on whether the separation step is or is not required, immunoassay methods can be classified into a het- erogeneous or a homogeneous assay, respectively. The analysis is ach- ieved by measuring the label activity in either of the bound or free fraction. On the other hand, the non-competitive assay, also called immunometric, uses an excess of labeled specific antibody toward the analyte of interest. It requires two antibodies that bind to non-overlapping epitopes on the analyte molecule. One of the two an- tibodies is bound to the solid phase, and the second one is labeled and used for detection. Labels commonly used in immunoassays include radioisotopes, fluorophores, and enzymes. For both types, the schematic diagrams are given below (for the competitive – Fig. 2, for the non-competitive – Fig. 3), reprinted from Ref. [109].

• Radioimmunoassay RIA. Detection is based on radioactive labeling of antigen with unstable isotopes such as I-125, C-14, H-3.

• Enzyme immunoassay EIA with sub-categories: ELISA – Enzyme- linked immunosorbent assay; EMIT – Enzyme-multiplied immuno- assay technique; MEIA – Microparticle enzyme immunoassay.

• Fluoroimmunoassay FIA

• Chemiluminescence immunoassay CLIA

• Immunonephelometry IN. Detection is based on an antigen-antibody complex scattering light.

• Immunoturbidimetry IT. Detection is based on an antigen-antibody complex blocking light thus increasing the turbidity of the sample.

As it was mentioned in the introductory part, the concentration of albumin higher than 2 mg/L could be indicative of potential kidney diseases [10]. Subsequently for many immunoassays methods for albu- min determination in urine the limit of detection (LOD) of magnitude around 1 mg/L would be enough for clinical purposes, thus priority should be given to the parameters such as specificity, accuracy, costs, etc. However for low abundant proteins achieving lower LOD is important [110]. In 1980–1990 common immunoassays for urinary proteins for clinical purposes were developed. Comper et al. compared four immunoassays (RIA, IN, two IT) and high-performance liquid chromatography (HPLC) [79]. The correlation coefficients calculated were high (>0.85) which is indicative of a strong linear relationship between all assays studied. While the correlation was high, the slopes of the regression line were significantly different from 1, it means that certain methods tend to underestimate or overestimate. For example, the authors found that the Beckman immunonephelometry assay consistently gives values approximately threefold lower than the Dade-Behring immunoturbidimetry assay. There have been numerous other studies that have found similar levels of variation between different immunoassays and other techniques, such as in the paper by Shaikh et al. [111]. The authors claim that theirs LC-MS and HPLC as- says both performed poorly at concentrations below 20 mg/L, while immunoturbidimetric assay underestimated albumin compared with LC-MS. Hoofnagle et al. state that immunoassays have four common problems of which a researcher should be aware [112]. Assay performed with one set of reagents does not measure the same concentration as another which results in patients having measurements performed on the same platform for continuity of care. The other three problems re- sults in falsely low or high results depending on the analyte concentra- tion, antibodies used. While immunoassays methods have their shortcomings they are still widely used in clinical practice. Modern immunoassays are actively being developed and some of their analytical parameters are given in the summary Table 3.

(8)

3.3. Fluorescence spectroscopy

Fluorescence is a luminescence phenomenon in which a compound emits light after absorption of electromagnetic irradiation [133]. In most cases, the emitted light has a longer wavelength, and therefore lower energy, than the absorbed radiation. Often in chemical analysis, a sample is irradiated in UV or visible range, and as a result, fluorescence occurs in the visible and near-infrared region respectively. Fluorescent materials cease to glow nearly immediately when the radiation source stops, unlike phosphorescent materials, which continue to emit light for some time after. Fluorescence spectroscopy is one of the most common techniques with applications in geology, chemistry, medicine, and as- tronomy [133]. Because of the sensitivity that the method affords, fluorescent molecule concentrations as low as 1 part per trillion can be measured [134]. Fluorescence spectroscopy can be used on its own, and also it is often used in conjunction with high-performance liquid

chromatography as a detector. Often fluorescence detection is used in immunoassays to either label antigen or antibody.

Usually, analyte molecules do not emit strong luminescence for their analytical detection. Hence an analyte usually modified with a fluo- rophore, or a fluorescent probe, in other words. The fluorophore (or a fluorescent probe) is a fluorescent chemical compound that can re-emit light upon light excitation. A suitable fluorescent probe should have the following characteristics: (i) is conveniently excitable, without simul- taneous excitation of the biological matrix, and detectable with con- ventional instrumentation; (ii) is bright, that is, possesses a high molar absorption coefficient at the excitation wavelength and a high fluores- cence quantum yield, (iii) is soluble in relevant buffers, cell culture media or body fluids, (iv) is sufficiently stable under relevant conditions, (v) has functional groups for site-specific labeling, (vi) has reported data about its photophysics, and (vi) is available in reproducible quality [135]. Most common fluorophores are usually organic molecules, that Fig. 2.Schematic diagram for the competitive immunoassays. Reprinted from Ref. [109].

(9)

typically contain several combined aromatic groups or planar or cyclic molecules with several π bonds. Fluorophores are sometimes used alone, as a tracer in fluids, as a dye for staining of certain structures, as a substrate of enzymes, or as a probe or indicator (when its fluorescence is affected by environmental aspects such as polarity or ions). More generally they are covalently bonded to a macromolecule, serving as a marker (or dye, or tag, or reporter) for affine or bioactive reagents (antibodies, peptides, nucleic acids). Fluorophores are notably used to stain tissues, cells, or materials in a variety of analytical methods, i.e., fluorescent imaging and spectroscopy. Some examples of widely used fluorophores are fluorescein, and by its amine reactive isothiocyanate derivative fluorescein isothiocyanate, methylene blue. Over the last two decades, the role of semiconductor quantum dots (QD) and other types of nanoparticles have grown considerably in bionanalysis and bio- imaging. Petryayeva et al. conclude that QDs as fluorophores and for other bioimaging purposes offer nontrivial advantages, such as the brightness needed for sensitive detection, the photostability needed for tracking dynamic processes, the multiplexing capability needed to elucidate complex systems or the nanoscale interface needed for bio- molecular engineering of novel probes and biosensors [136]. The fluo- rescent labeling in immunoassays is the common practice (see page 20).

A method of fluorescence, where two light-sensitive molecules, that transfer energy between them, is called fluorescence resonance energy transfer. Fluorescence resonance energy transfer (FRET) involves the transfer of energy from a fluorescent donor molecule via a nonradiative dipole-dipole interaction to an acceptor molecule. The emission of the energy transfer-excited acceptor can be distinguished from donor emission by the use of a longer-wavelength filter (spatial resolution).

The fluorescent lanthanide chelates of europium and terbium are attractive FRET donors. The use of a long-lifetime donor fluorophore and a short-lifetime acceptor fluorophore together with pulsed-laser excitation and time-resolved detection (temporal resolution) are effec- tive in reducing the background signal. Qin et al. used time-resolved FRET for the determination of urinary albumin [118]. Particularly, they used, in the competitive homogeneous assay, an albumin-specific monoclonal antibody labeled with a stable fluorescent europium chelate as donor and an albumin labeled with cyanine 5 (Cy5) as acceptor. The limit of detection was 5.5 mg/L with the detectable range

being 10–320 mg/L. Another interesting approach to fluorescence spectroscopy is the fluorescent quenching. Chawjiraphan et al. devel- oped a method to determine HSA in urine and serum by the graphene oxide-mediated (GO) fluorescence quenching aptasensor [132]. When albumin was added to the complex GO with the fluorescence-labeled aptamer, the aptamer detached from the complex to bind albumin, which resulted by an increase in fluorescence intensity, as shown in Fig. 4. The limit of detection was 0.05 mg/L and the detection range is 0.1–14.0 mg/L. Other quantitative works using the fluorescence spec- troscopy are summarized in Table 4.

3.4. The Infrared and Raman spectroscopies

The absorption of infrared radiation IR excites vibrational transitions of molecules [148]. The infrared spectral region covers wavelengths from 780 nm to 1000 μm which can be further subdivided into the near-infrared region from 780 nm to 2500 nm, the mid-infrared region from 2500 nm to 50 μm and the far-infrared region from 50 μm to 1000 μm. Since vibrational frequency and probability of absorption depend on the strength and polarity of the vibrating bonds, they are influenced by intra- and intermolecular effects [149]. The approximate position of an infrared absorption band is determined by the vibrating masses and the type of bond (single, double, triple), the exact position by electron-withdrawing or donating effects of the intra- and intermolec- ular environment and by coupling with other vibrations. Information that can be derived from the infrared spectrum:

• Chemical structure of the vibrating group

• Chemical properties of neighbouring groups in a molecule

• Redox state

• Bond parameters

• Hydrogen bonding

• Electric fields

• Conformational freedom

Basically, all polar bonds contribute to the infrared absorption so there is no need to specifically label biomolecules to detect them.

However, the infrared spectrum of biomolecules is challenging to Fig. 3. Schematic diagram for the non-competitive immunoassay. Reprinted from Ref. [109].

(10)

Table 3

Summary table of quantitative methods for application of immunoassay for protein determination in the human urine matrix (unless otherwise stated in the method section). CV (coefficient of variation) parameters are taken from maximum readings, for Recovery – the range is given.

Method Reagent for detection (if any) Analyte The limit of

detection, LOD

Range Analytical and

experimental parameters

Reference

Fluoroimmunoassay Fluorescein isothiocyanate HSA 0.5 mg/L 0.520 mg/L (linear R2 =

0.99) Total time for

experiment 4–6 h Recovery 105 ±7%

19 Chavers et al., 1984 [113]

Immunoturbidimetry No reagent HSA 4 mg/L 4–35.5 mg/L (linear) Time so screen 15 min

Recovery 95%–104% 20 Lloyd et al., 1987 [114]

Immunonephelometry No reagent HSA 0.34 mg/L 0.34–43.0 mg/L

(detectable) Automated version 240 samples an hour CV (intraassay) 6% CV (interassay) 9%

21 Marre et al., 1987 [115]

Immunoturbidimetry No reagent HSA 2 mg/L 2–260 mg/L (detectable) CV (interassay) 4.8% 22 Bakker et al.,

1988 [116]

Chemiluminescence Acridinium ester HSA 0.01 mg/L 0.3–10 mg/L (detectable) Recovery 90–106% 23 Horton et al., 1989 [117]

Radioimmunoassay Na125I HSA 0.015 mg/L 1.1814.96 mg/L

(detectable) Recovery 90106% 24 Horton et al., 1989 [117]

Time resolved Fluorescence

Resonance Energy Transfer Europium chelate (donor)/

cyanine 5 (acceptor) HSA 5.5 mg/L 10–320 mg/L (detectable) Experiment time 12 min Recovery 103–122% CV (within-run) 6.9–10%

CV (between-run) 7.5–13%

25 Qin et al., 2003 [118]

Chemiluminescence enzyme- linked immunosorbent assay based on Avidin–biotin

4-methoxy4-(3- phosphatephenyl)-spiro- (1,2-dioxetane-3,2 -adamantane) (AMPPD) to detect enzyme activity

HSA 0.089 mg/L 0.1515 mg/L (linear R2

=0.9902) Low usage of antibody CV (intra-assay) 10.7%

CV (inter-assay) 15%

Recovery 112%

26 Zhao et al., 2005 [119]

Chemiluminescence enzyme- linked immunosorbent assay based on fluorescein- isothiocyanate (FITC)

4-methoxy4-(3- phosphatephenyl)-spiro- (1,2-dioxetane-3,2 -adamantane) (AMPPD) to detect enzyme activity

HSA 0.089 mg/L 0.15–15 mg/L (linear R2

=0.9906 CV (intra-assay) 10.5%

CV (inter-assay) 11.9%

Recovery 109%

27 Zhao et al., 2005 [119]

Magnetic two-site

immunoassay Dextran-coated nanoscaled to detect magnetic permeability

HSA 5 mg/L 0–400 mg/L (linear) Experiment time 6.5

min CV 11% 28 Lu et al., 2006 [120]

Capillary electrophoresis immunoassay with spectrophotometric detection

Hydroxylamine HSA 0.60 mg/L 06.67 mg/L (detectable) Voltage 12 kV Recovey

91–160% 29 Giovannoli

et al., 2007 [66]

Electrochemical

immunosensor Fe3O4/Au colloid -modified

gold electrode NMP22 Nuclear Matrix Protein 22

5⋅104 mg/

L 1.2⋅103-0.2 mg/L

(linear, R2 =0.9932) CV 3.9% 30 Ning et al., 2007 [121]

Resonance scattering (RS)

spectral immunoassay Cu2O for RS HSA 7.2106

mg/L 1.41054.3104 mg/L CV 4.5% Recovery

92.1–106.8% 31 Jiang et al., 2009 [122]

Fluoroimmunoassay, a quantum-dot (QS) optical immunosensor (the buffer matrix)

CdSe/ZnS QS HSA 0.032 mg/L 0.2–200 mg/L

(detectable) Low cost

immunosensor. 32 Tu et al., 2012 [123]

Time-resolved fluorescence resonance energy transfer immunoassay

Europium based oligonucleotides (donor)/

oligonucleotide(acceptor)

HSA 0.0039 mg/

L 0.0039–0.5 mg/L (linear) 0.00391 mg/L (detectable)

Experiment time 40 min Tested in serum, salive, urine CV 3%

33 Wang et al., 2012 [124]

Chemiluminescence lateral

flow immunoassay Luminol/enhancer/H2O2

with silicon photosensor HSA 2.5 mg/L 2.5–850 mg/L

(detectable) CV 20% Recovery

85–112% 34 Zangheri

et al., 2016 [125]

Amperometric immunosensor

(the buffer matrix) Carbon nanotubes/gold nanoparticles screenprinted electrodes

p53 protein 6.1⋅104

mg/L 8.4⋅104-0.437 mg/L

(linear) CV 10% Recovery

84–144% 35 Giannetto

et al., 2017 [126]

Piezoelectric immunosensor,

label-free immunoassay Quartz crystal coated with a

silver electrode HSA 0.0095 mg/

L 0.1–100 mg/L Diagnostic sensitivity

98.7% Diagnostic specificity 100%

36 Theansun et al., 2017 [127]

Time-resolved fluorescence

immunoassay Europium (III) and samarium

(III) cystatin-C 1.26 103

mg/L 0.42 103-0.9563 mg/L

(linear, R2 =0.9991) Recovery 99.36% CV 3.2–5.9% (inter) CV 3.46.3% (intra)

37 Liu et al., 2017 [128]

Time-resolved fluorescence

immunoassay Europium (III) and samarium

(III) β2-microglobulin 2.13 103

mg/L 0.86 103-0.9754 mg/L

(linear, R2 =0.9998) Recovery 100.18% CV 3.0–6.8% (inter) CV 3.1–8.4% (intra)

38 Liu et al., 2017 [128]

Chemiluminescence

immunoassay Magnetic microparticles using carboxylic acid groups, acridinium

monomeric

laminin-γ2 1.0 105

mg/L 1.0 105-0.02 mg/L

(detectable) Recovery 82.3–96.2% 39 Nakagawa et al., 2017 [129]

Fluoroimmunoassay, optical

microchips Streptavidin-fusion based

variants HSA 0.65 mg/L Limit of quantification

2 mg/L 40 Semeradtova

et al., 2018 [130]

Electrochemical ELISA 1 103-0.01 mg/L (linear)

(continued on next page)